Health product recall

BABYTHERM 8004/8010 (2016-06-27)

Starting date:
June 27, 2016
Posting date:
July 22, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-59464

Affected products

A. BABYTHERM 8004/8010

Reason

Dragerwerk AG & Co. KGaA, the parent company of Drager Medical Canada Inc., became aware of a case in which a Babytherm 8004 open care warmer was operated in skin temperature mode with skin temperature sensor to prepare it for use. There was no patient in the warmer at the time. In skin temperature mode, the radiant warmer output is controlled automatically until the set skin temperature is reached. As a result of operating the temperature sensor without patient contact, the radiant warmer was operated continuously at a high output level. During this forced continuous operation under high thermal load, the heating element aged faster. As a result, a piece of the heating element detached and fell onto the mattress tray where a towel ignited. There were no reported injuries.

Affected products

A. BABYTHERM 8004/8010

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

FR 00098

FR 00114

FR 00120

Companies
Manufacturer

DRAEGERWERK AG & CO

MOISLINGER ALLEE 53-55

LUBECK

23542

GERMANY