B-type Natriuretic Peptide Test Kits (2017-01-18)
- Starting date:
- January 18, 2017
- Posting date:
- February 9, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-62120
Affected Products
- B-type Natriuretic Peptide (BNP) 100 Test Kit
- B-type Natriuretic Peptide (BNP) 500 Test Kit
Reason
L'enquête de Siemens concernant les plaintes a confirmé que le paramètre d'étalonnage « faible écart de l’étalonneur » peut être inférieur à la plage acceptable, ce qui entraîne un statut d'étalonnage invalide et empêche les clients de générer des résultats pour le dosage des peptides natriurétiques de type B (BNP). Lorsque l’étalonnage et le contrôle qualité (QC) sont valides, les résultats des patients sont valables et acceptables pour être présentés. Lorsqu’on n’obtient pas un étalonnage valide, le contrôle de la qualité et le test du patient ne peuvent être effectués.
Affected products
A. B-type Natriuretic Peptide (BNP) 100 Test Kit
Lot or serial number
Lots ending in 186, 188, 189 and 191
Model or catalog number
- 02816138
- 02816634
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
B. B-type Natriuretic Peptide (BNP) 500 Test Kit
Lot or serial number
Lots ending in 186, 188, 189, 191
Model or catalog number
- 02816138
- 02816634
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES