Azurion 7 M20
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Azurion 7 M20 |
All lots. |
722 224 |
Issue
Philips has become aware of a potential safety issue with the Azurion 7 M20 Flexarm System ("flexarm") with a Maquet Magnus Table with software versions 05.02.12 and 05.02.15 ("table"). Due to a communication issue between the flexarm and the table, motorized movements of the table and/or the flexarm may be lost. A representative of Getinge (manufacturer of the Maquet Magnus Table) will arrange a site visit to return the software of your table to version 05.02.12. After this change, your system will be subject to the medical device recall letter (HS23039R/2023-IGT-BST-020).
Recall start date: December 13, 2023