This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Axsym System - Valproic Acid Assay (2013-10-31)

Starting date:
October 31, 2013
Posting date:
December 4, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-36983

Recalled products

A. Axsym System - Valproic Acid Assay 

Reason

Abbott has confirmed that some kits of Axsym Valproic Acid Reagent lot 18439M503 have the potential to generate invalid calibration, control or sample runs due to net intensity too low, resulting in axsym instrument error code 1072. If the error occurs during calibration, the curve will be invalid and recalibration will be required. If the error occurs during a control/sample run, no results will be reported.

Affected products

A. Axsym System - Valproic Acid Assay 

Lot or serial number
  • 18439M503
Model or catalog number
  • 7A71-20
Companies
Manufacturer
Abbott Laboratories Diagnostcis Division
100 Abbott Park Road,
Abbot Park
60064
Illinois
UNITED STATES