This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Axiom Vertix MD Trauma (2014-03-05)

Starting date:
March 5, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38681

Recalled Products

  1. Axiom Vertix MD Trauma

Reason

In rare cases, steel cables inside the lifting column of the system can fail without triggering the safety lock. This can result in the system U-arm dropping down unexpectedly during movement in the vertical direction. Since the U-arm weight is approximately 100 kg, failure of the steel cables may result in serious injury to patients.

Affected products

A. Axiom Vertix MD Trauma

Lot or serial number

Serial numbers:

  • 1041
  • 1058
Model or catalog number
  • 8908290
Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY