This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
AW Volumeshare - Seno Advantage Software Option
- Starting date:
- September 18, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Assistive/Medical Devices
- Audience:
- General Public (GP), Healthcare Professionals, Hospitals
- Identification number:
- RA-15123
Reason
GE Healthcare has become aware that patient mismatch without a warning message could occur on Seno Advantage (SA) Workstations under the following circumstances:
1. New non-mammographic images for patient a are being received (for example, via the network) at the same time that mammographic images for patient a are displayed on the high-resolution black and white monitors.
2. The non-mammographic images are then selected and displayed on the color monitor.
3. Mammographic images from a different patient are then selected on the high-resolution black and white monitors.
Affected products
AW Volumeshare - Seno Advantage Software Option
Lot or serial number
> 10 Numbers, Contact manufacturer
Model or catalog number
5183685-X, 5183859-X
Companies
- Manufacturer
-
GE Medical Systems SCS
AW Volumeshare - Seno Advantage Software Option
Lot or serial number
'00000193277GE1, '000004410002GS
Model or catalog number
5341143-X, 5342425-X
Companies
- Manufacturer
-
GE Medical Systems SCS