Health product recall

Auto Suture Blunt Tip and Structural Balloon Trocar

Brand(s)
Last updated

Summary

Product
Auto Suture Blunt Tip and Structural Balloon Trocar
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalog number

Auto Suture Blunt Tip Trocar

More than 10 numbers, contact manufacturer.

OMS-T10BTS
OMS-T10BTNL
OMS-T10BTSNL
OMS-T12BT
OMS-T10BT
OMS-T12BTNL

Auto Suture Structural Balloon Trocar

More than 10 numbers, contact manufacturer.

OMS-T10SB

Issue

Medtronic received reports of events related to Covidien Auto Suture™ Structural Balloon Trocar and Auto Suture™ Blunt Tip Trocar reporting trocar seal disengagement when using the trocar with certain mesh products. This issue can occur when the trocar is used to deploy the mesh not in accordance with the mesh instructions for use (IFU). Not following the IFU for the mesh including use of incompatible trocar size, excess force, and not hydrating the mesh and/or folding of the mesh could result in a seal disengaging into the cavity.

Recall Start Date: March 4, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Gastroenterology and urology
Health products - Medical devices - General and plastic surgery
Companies

Covidien Llc

15 Hampshire Street, Mansfield, Massachusetts, United States, 02048

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-75231

Get notified

Receive notifications for new and updated recalls and alerts by category.

Subscribe