Atellica IM 1300/1600 Analyzer (2021-05-20)
- Starting date:
- May 20, 2021
- Posting date:
- June 10, 2021
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-75779
Last updated:
Affected Products
A. Atellica IM 1300 analyzer
B. Atellica IM 1600 analyzer
Reason
Siemens has identified a set of valves that are used on the Atellica CH 930, Atellica IM 1300 and Atellica IM 1600 analyzers that may develop a malfunction due to a manufacturing defect and may result in the valve wearing and leaking over time. The leaking is a risk in valves subject to high pressure or high wear and may impact the result accuracy of any assay.
Affected products
A. Atellica IM 1300 analyzer
Lot or serial number
More than 10 numbers, contact manufacturer.
Model or catalog number
11066001
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES
B. Atellica IM 1600 analyzer
Lot or serial number
IH00244
IH00255
IH00259
IH01519
IH01520
IRH008672002
Model or catalog number
11066000
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
UNITED STATES