Health product recall

Assay Tip/Assay Cup (2019-08-15)

Starting date:
August 15, 2019
Posting date:
August 30, 2019
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-70845


Last updated:

Affected Products

Assay Tip/Assay Cup

Reason

Assaytips were identified with abnormal internal structure in some lots. The assaytips are part of the assaytip / assaycup tray (product no. 05694302001). Recently, Roche has received a small number of customer complaints regarding defective assaytips. The affected assaytip / assaycup tray lots have been distributed from August 2018 until July 2019 (list of affected lots above). Internal investigation shows that the average amount of potentially affected assaytips in those lots is approximately 1.5%. The overall incidence rate represents 0.14% Of the assaytips production since the introduction of the cobas e 801.

Affected products

Assay Tip/Assay Cup

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

5694302001

Companies
Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY