Health product recall

ASP200S/ASP300S TISSUE PROCESSOR(2015-08-05)

Starting date:
August 5, 2015
Posting date:
August 24, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-54762

Affected Products

A. ASP200S/ASP300S TISSUE PROCESSOR

Reason

One complaint was received where the remote alarm of an affected instrument did not alert the customer properly of an error that occurred during processing. A review meeting was held by the manufacturer to determine that the drawing of the remote alarm cable does have an error which led to a malfunction of the remote alarm. Cables with this error were used since April 24, 2013 in production. A ship hold of the ASP200s/ASP300s was initiated on July 13, 2015. Within this time period 901 instruments were distributed globally with the affected cables. The affected devices have an incorrect internal wiring to the remote alarm plug, which can cause the remote alarm to not function correctly in the case of an instrument failure.

Affected products

A. ASP200S/ASP300S TISSUE PROCESSOR

Lot or serial number

5541

5593

5618

5685

5686

5687

5818

Model or catalog number

ASP200S

ASP300S

Companies
Manufacturer

Leica Microsystems Nussloch GMBH

HEIDELBERGER STREET 17-19

Nussloch

69222

GERMANY