This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Artis Zee Components (2016-03-07)

Starting date:
March 7, 2016
Posting date:
March 17, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-57576

Affected Products

A. ARTIS ZEEGO MULTI-AXIS

B. ARTIS ZEE MULTI-PURPOSE

C. ARTIS ZEE FLOOR

D. ARTIS ZEE CEILING

E. ARTIS ZEE BIPLANE

Reason

In the framework of the regular post market surveillance the following problems have been identified:  A. In Artis zeego systems, angulations in the vicinity of the c-arm collision area can cause the cable inlet to become clamped and can result in mechanical damage.  B. In Artis systems with A100Plus generators, an attempt to resume operation following detection of a fault (such as a short circuit in the x-ray tube) can result in the failure of a module in the high-voltage generator.

Affected products

A. ARTIS ZEEGO MULTI-AXIS

Lot or serial number

160109

160354

160810

160846

160936

161003

161020

Model or catalog number

10280959

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



B. ARTIS ZEE MULTI-PURPOSE

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

10094139

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



C. ARTIS ZEE FLOOR

Lot or serial number

136972

137078

137167

137344

137423

137457

137561

Model or catalog number

10094135

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



D. ARTIS ZEE CEILING

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

10094137

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY



E. ARTIS ZEE BIPLANE

Lot or serial number

154569

154679

154691

154693

Model or catalog number

10094141

Companies
Manufacturer

Siemens AG

Wittelsbacherplatz 2

Muenchen

80333

GERMANY