Health product recall

Artis Q, Icono and Pheno Systems

Last updated

Summary

Product
Artis Q, Icono and Pheno Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products Lot or serial number Model or catalogue number
Artis Q More than 10 numbers, contact manufacturer. 10848281 10848283 10848280 10848282
Artis Q.Zen 123214 123506 105103 10848355 10848353
Artis Icono Floor 170369 170382 11327700
Artis Pheno 164123 164367 10849000
Artis Icono Biplane 180441 11327600

Issue

The Artis system is provided with a dedicated error detection mechanism. In rare cases of failure of this error detection mechanism it will not be possible to release x-ray anymore until system shutdown.

Recall start date: August 12, 2022

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Siemens Healthcare Gmbh

Henkestr. 127, Erlangen, Germany, 91052

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-64558

Get notified