ARTIS Pheno
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
| Affected products | Lot or serial number | Model or catalogue number |
|---|---|---|
| ARTIS Pheno | 164123 | 10849000 |
| ARTIS Pheno | 164101 | 10849000 |
Issue
Initiating a shutdown of the system can sporadically lead to a state in which the robotic C-arm stand cannot be moved after the next system startup. Other movements (e.g. table movements) are still available with reduced speed.
The system may show the following generic messages "reduced stand/table speed" and "limited stand movement, wait".
In case the error occurs the C-arm stand will not move anymore. Therefore, the C-arm cannot be moved into the required position. This may result in a situation where it is necessary to cancel the clinical procedure or to begin or continue the procedure on an alternative system.
Recall start date: August 13, 2025
Additional information
Details
| Siemens Healthcare Gmbh |
| Henkestr. 127, Erlangen, Germany, 91052 |
Get notified
Receive emails about new and updated recall and safety alerts.