Artis Icono Ceiling
Brand(s)
Last updated
Summary
Product
Artis Icono Ceiling
Issue
Medical devices - Performance
What to do
Contact the manufacturer if you require additional information.
Audience
Healthcare
Affected products
Affected products | Lot or serial number | Model or catalogue number |
---|---|---|
Artis Icono Ceiling | 190027 190036 | 11328100 |
Issue
Artis Icono Ceiling systems are equipped with an automatic collision detection at locations where collisions with users or equipment might occur. If the issue occurs, the collision sensor around the x-ray tube housing shows the following unintended behavior: when the Artis Icono Ceiling system was switched off for a longer time (e.g. overnight) and switched on, this collision sensor is falsely signaling non-existing collisions for up to half an hour.
Recall start date: October 19, 2023
Additional information
Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies
Siemens Healthcare Gmbh |
Henkestr. 127, Erlangen, Germany, 91052 |
Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Identification number
RA-74549
Get notified
Receive notifications for new and updated recalls and alerts by category.