Health product recall

ARTIS DIALYSIS SYSTEM(2015-09-17)

Starting date:
September 17, 2015
Posting date:
October 8, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55296

Affected Products

SYSTÈME DE DIALYSE ARTIS

Reason

There have been reports of a burning smell/smoke coming out from the upper ventilation grids of the Artis dialysis machine. The events have been associated with a limited number of power supply units manufactured with a specific batch of capacitor. The technical analysis identified that the issue was caused by a defective capacitor inside the power supply unit which has the potential to melt after approximatively 4 years of use.

Affected products

ARTIS DIALYSIS SYSTEM

Lot or serial number

More than 100 numbers, contact manufacturer.

Model or catalog number

112949