This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Artis Devices (2016-12-14)

Starting date:
December 14, 2016
Posting date:
January 6, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-61708

Affected Products

A. Artis Zeego Multi-Axis

B. Artis Zee Multi-Purpose

C. Artis Zee Ceiling

D. Artis Zee Biplane

E. Artis Q

F. Artis Q.Zen

Reason

1. In Artis Systems with A100Plus or A100G Generators, an attempt to resume operation following detection of fault (such as a short circuit in the x-ray tube) can result in the failure of a module in the high-voltage generator.

2. For biplane systems delivered with software version VD11B since April of this year, software problems in conjunction with the graphics card may in rare cases result in the loss of image display in the examination room.

Affected products

A. Artis Zeego Multi-Axis

Lot or serial number

All lots

Model or catalog number

10280959

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY



B. Artis Zee Multi-Purpose

Lot or serial number

All lots

Model or catalog number

10094139

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY



C. Artis Zee Ceiling

Lot or serial number

All lots

Model or catalog number

10094137

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY



D. Artis Zee Biplane

Lot or serial number

All lots

Model or catalog number

10094141

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY



E. Artis Q

Lot or serial number

All lots

Model or catalog number

10848280

10848281

10848282

10848283

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY



F. Artis Q.Zen

Lot or serial number

All lots

Model or catalog number

10848353

10848355

Companies
Manufacturer

Siemens Healthcare GMBH

HENKESTRASSE 127

ERLANGEN

91052

GERMANY