This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Artas Disposable Kit (2013-10-14)

Starting date:
October 14, 2013
Posting date:
December 17, 2013
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-37211

Recalled products

A. Artas Disposable Kit 

Reason

There is a discoloration on the tensioner pins that was observed prior to performing the Artas procedure.  This discoloration has been identified as a compound of iron and chlorine and found to be a remnant of the cleaning process used on the stainless steel (SS303) tensioner pins by the pin manufacturer.  Action has been taken to ensure that replacement product uses pins that have been properly cleaned and all potentially affected product in the field are to be returned and replaced.

Affected products

A. Artas Disposable Kit 

Lot or serial number
  • 12490
Model or catalog number
  • FP-45384
  • FP-45616
Companies
Manufacturer

Restoration Robotics

128 Baytech Drive

San Jose

95134

California

UNITED STATES