Health product recall

ARES CATHETER (2020-12-08)



Starting date:
December 8, 2020
Posting date:
December 24, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74649


Last updated:

Affected Products 

A. ARES VENTRICULAR CATHETER WITH RIGHT ANGLE CLIP AND STYLE

B. ARES PERITONEAL CATHETER

C. ARES KIT VENTRICULAR CATHETER WITH RIGHT ANGLE CLIP AND STYLET AND PERITONEAL CATHETER

Reason

During routine post-sterilization inspection, Medtronic identified that there is a potential for a defect on the seal of the outer pouch of specific lots of the Ares catheters. Internal testing has shown that approximately 3 percent of pouches are potentially impacted by this condition. This defect may compromise the sterility of the pouch contents, which may increase the risk of post-operative infection, requiring further medical intervention. To date, Medtronic has not received any reports of patient harm or any complaints related to this issue.

Affected products

A. ARES VENTRICULAR CATHETER WITH RIGHT ANGLE CLIP AND STYLE

Lot or serial number

10265137

10265138

10278409

10297691

10297693

10297696

10353501

Model or catalog number

91101

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



B. ARES PERITONEAL CATHETER

Lot or serial number

10273893

10278418

10312362

10316961

Model or catalog number

93092

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES



C. ARES KIT VENTRICULAR CATHETER WITH RIGHT ANGLE CLIP AND STYLET AND PERITONEAL CATHETER

Lot or serial number

More than 10 numbers, contact manufacturer. 

Model or catalog number

95001

Companies
Manufacturer

Medtronic Inc.

710 Medtronic Parkway

Minneapolis

55432

Minnesota

UNITED STATES