This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health professional risk communication

Archived - Infections from Endoscopes Inadequately Reprocessed by Automatic Endoscope Reprocessing Units

Starting date:
June 30, 2000
Posting date:
June 30, 2000
Type of communication:
Notice to Hospitals
Subcategory:
Medical Device
Source of recall:
Health Canada
Audience:
Healthcare Professionals
Identification number:
RA-1700070

Notice about Health Canada advisories

Medical Devices Alert 112

June 30, 2000

To: Hospital Administrators

Attention:

  • Hospital Risk Management
  • Infection Control
  • Surgical Services and Operating Room
  • Ambulatory Surgical Centres
  • Practitioners who use automated endoscope reprocessors

Subject: Infections from Endoscopes Inadequately Reprocessed by Automatic Endoscope Reprocessing Units

The purpose of this Alert is to advise you of the potential for serious infections from endoscopes that are inadequately decontaminated by an automatic endoscope reprocessing unit (AERU) and to provide recommendations for reducing the risk of infection.

The US Centers for Disease Control and Prevention and the US FDA recently published reports Footnote 1,Footnote 2 on cases of apparent patient-to-patient infections following procedures with bronchoscopes that had been inadequately reprocessed by AERUs. The reports identified inconsistencies between the reprocessing instructions provided by the AERU manufacturer and the endoscope manufacturer and found that bronchoscopes were inadequately reprocessed when inappropriate channel connectors were used with the AERU.

The Health Protection Branch is also aware of instances where health care facilities have identified growth of pathogenic organisms in critical components of the AERU even when recommended AERU maintenance had been followed.

Recommendations

The Health Protection Branch recommends that health care facilities that reprocess endoscopes take the following precautions.

A. Choosing an AERU

Before you purchase an AERU, make sure that it meets the following criteria.

  • the AERU is licensed for sale in Canada
  • there are no potential reservoirs of infections - areas in the AERU where water or disinfectant can stagnate.
  • the AERU can effectively irrigate all channels of any endoscope to be processed in the AERU, ensuring effective contact between the chemical disinfectant and the channel walls.
  • all fluids (water and chemicals) are completely drained and discarded following each cycle in the processing period, ensuring that reuse of contaminated fluid is not possible.
  • the cycle length can be adjusted to correspond with the type of endoscope and its intended use, to ensure high-level disinfection or sterilization.
  • the manufacturer of the AERU identifies by brand and model each endoscope that may be effectively reprocessed in the AERU and the limitations of the AERU in processing certain models of endoscopes and accessories.

B. Cleaning and Reprocessing Procedures

  1. Ensure that the cleaning and reprocessing procedures used by the staff are current and effective. Use the guidelines developed by the Canadian Society of Gastroenterology Nurses Association Footnote 3 and the guidelines of the Society of Gastroenterology Nurses AssociationFootnote 4 to evaluate the reprocessing recommendations received from AERU manufacturers and endoscope manufacturers.
  2. Ensure that all staff who handle soiled endoscopes are fully trained to handle and clean endoscopes in accordance with the endoscope manufacturer's specific instructions for each brand and model. Ensure that the written procedures are being followed exactly.
  3. The following steps are critical regardless of whether your facility manually reprocesses the endoscopes or uses an AERU.
    • Flush the endoscope immediately after every procedure.
    • Before any automated or manual disinfection or sterilization procedure:
      1. clean the endoscope manually, meticulously removing all organic material in or on the endoscope; and
      2. rinse all parts and lumens thoroughly with copious amounts of water and dry them before placing the scope in the reprocessing unit. This drying step avoids dilution of the disinfectant concentration or interaction between the water and the sterilant.
  4. Follow your endoscope manufacturer's directions for reprocessing that specific brand and model. Check with the manufacturer of the endoscope to determine that the specific brands and models of endoscopes you use can be reprocessed in an AERU and whether specific steps are required for automated reprocessing. Not all endoscopes can be reprocessed using an AERU.
  5. If some brands and models of endoscopes used in your facility are not specifically indicated in the AERU labelling, ask the manufacturer of the AERU if these endoscopes have been tested with the AERU.
  6. Be sure to reconcile any differences in processing recommendations between the endoscope manufacturer, the AERU manufacturer and the manufacturer of the disinfectant, particularly with respect to the instructions for the use of channel connectors or capping of specific channels.
  7. In the absence of detailed written instructions for processing the specific model of endoscopic, the endoscope must be reprocessed manually.
  8. Regardless of whether you manually reprocess or use an AERU, incorporate a final drying step in you reprocessing protocol; flush all channels with alcohol, followed by purging of channels with air to remove the alcohol (American Society for Testing Materials Standard F1518-94).
  9. Implement a comprehensive quality assurance program. Your QA program should include:
    1. Visual inspection of the reprocessing procedure and reprocessed endoscopes to identify conditions that may affect cleaning and disinfection processes.
    2. Assurance that all manufacturers' recommended maintenance schedules for endoscopes and for AERUs are performed.
    3. Use of appropriate process monitors for the disinfectant in the AERU or in manual basins.
    4. Records of the use of each endoscope, showing the patient upon which it was used, the type of procedure and the reprocessing process (unit) used.
    5. Regular review of the training given to reprocessing Additional staff straining and supervised practice each time a new endoscopy model is introduced into the facility.
    6. A surveillance system capable of detecting clusters of infections or pseudo-infections associated with endoscopic procedures.

Reporting Adverse Events

Reports of adverse incidents involving the use of endoscopes and AERUs and inquiries concerning this Alert should be addressed to:

The Director, Medical Devices Bureau, Therapeutic Products Programme, Room 1605, Main Statistics Canada Building, Tunney's Pasture, Address Locator 0301H1, Ottawa, Ontario, K1A 0L2. Tel: 613-957-4786, fax: 613-957-7318.

Diane Gorman

Assistant Deputy Minister

References:

Footnote 1

Centers for Disease Control and Prevention. Bronchoscope-related infections and pseudoinfections -New York, 1996 and 1998. MMWR 1999; 48(26); 557-560.

Return to footnote 1 referrer

Footnote 2

FDA and CDC Public Health Advisory: Infections from endoscopes inadequately reprocessed by an automated endoscope reprocessing system, September 10, 1999.

Return to footnote 2 referrer

Footnote 3

Canadian Society of Gastroenterology Nurses Association. Infection Control: Recommended Guidelines in Endoscopy settings. Ottawa, 1998.

Return to footnote 3 referrer

Footnote 4

Society of Gastroenterology Nurses Association: Recommended Guidelines for Infection Control (2nd ed.). Rochester NY: 1992.

Return to footnote 4 referrer