This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Architect I System - EBV VCA IgM controls (2014-05-14)

Starting date:
May 14, 2014
Posting date:
June 5, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-39825

Recalled Products

Architect I System - EBV VCA IgM controls 

Reason

During stability testing a significant signal drop for the positive control of architect EBV VCA IgM control kit compared to the release results was recognized, indicating a loss in signal over time. Subsequent out of specification low results for the positive control of Architect EBV VCA IgM control kit was documented confirming above observation of a loss in signal over time.

Affected products

Architect I System - EBV VCA IgM controls 

Lot or serial number

38938LI00

Model or catalog number

3P66-10

 

Companies
Manufacturer

Abbott GmbH and Co. K.G.

Max Planck Ring 2

Wiesbaden

65205

GERMANY