Health product recall

ARCHITECT I SYSTEM - EBV VCA IgM Calibrator (2020-11-03)



Starting date:
November 3, 2020
Posting date:
November 13, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-74305


Last updated:

Affected Products

ARCHITECT I SYSTEM -  EBV VCA IgM Calibrator

Reason

Abbott has identified the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 may show reduced RLU (Relative Light Units) signal. This reduced RLU signal may cause a decrease in cutoff RLU values in assay calibration resulting in:

  • upward shift in quality control S/CO values
  • non-reactive to false grayzone patient results in the range of 0.50 - 0.56 S/CO
  • grayzone to false reactive patient results in the range of 1.00 - 1.13 S/CO

The effect of decreased cutoff RLU values is observed when using the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 and is independent from the ARCHITECT EBV VCA IgM Reagent lot used.

Affected products

ARCHITECT I SYSTEM -  EBV VCA IgM Calibrator

Lot or serial number

08159BE00

Model or catalog number

3P66-01

Companies
Manufacturer

Abbott GmbH

Max Planck Ring 2

Wiesbaden

65205

GERMANY