ARCHITECT I SYSTEM - EBV VCA IgM Calibrator (2020-11-03)
- Starting date:
- November 3, 2020
- Posting date:
- November 13, 2020
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-74305
Last updated:
2020-11-13
Affected Products
ARCHITECT I SYSTEM - EBV VCA IgM Calibrator
Reason
Abbott has identified the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 may show reduced RLU (Relative Light Units) signal. This reduced RLU signal may cause a decrease in cutoff RLU values in assay calibration resulting in:
- upward shift in quality control S/CO values
- non-reactive to false grayzone patient results in the range of 0.50 - 0.56 S/CO
- grayzone to false reactive patient results in the range of 1.00 - 1.13 S/CO
The effect of decreased cutoff RLU values is observed when using the ARCHITECT EBV VCA IgM Calibrator lot 08159BE00 and is independent from the ARCHITECT EBV VCA IgM Reagent lot used.
Affected products
ARCHITECT I SYSTEM - EBV VCA IgM Calibrator
Lot or serial number
08159BE00
Model or catalog number
3P66-01
Companies
- Manufacturer
-
Abbott GmbH
Max Planck Ring 2
Wiesbaden
65205
GERMANY