This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Architect Free T4 Reagent

Starting date:
May 1, 2017
Posting date:
September 5, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64408

Reason

Recent data generated by Abbott have determined that the measuring interval is 0.40 ng/dl to 5.00 ng/dl. Measuring interval is defined as the range of values in ng/dl which meets the limits of acceptable performance for both imprecision and linearity. The current measuring interval for list numbers 7K65-29, 7K65-34 and 7K65-39 is 0.40 ng/dl to 6.00 ng/dl. A new FT4 assay file and updated package insert will be created reflecting the new measuring interval of 0.40 ng/dl to 5.00 ng/dl to be used with list numbers 7K65-29, 7K65-34 and 7K65-39.

Affected products

Architect Free T4 Reagent

Lot or serial number

More than 10 numbers, contact manufacturer

Model or catalog number

  • 7K65-29
  • 7K65-34
  • 7K65-39

Companies

Manufacturer

ABBOTT IRELAND DIAGNOSTICS DIVISION

LISNAMUCK, LONGFORD

IRELAND