This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Architect C16000/C4000 System - Processing Modules (2017-10-04)

Starting date:
October 4, 2017
Posting date:
October 13, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-64848

Affected products

  1. ARCHITECT C 16000 - PROCESSING MODULE
  2. ARCHITECT C4000 SYSTEM - PROCESSING MODULE

Reason

Abbott has become aware that the Peristaltic Head tubing placed on the Architect has the possibility to leak due to a manufacturing issue associated with the tubing connector. This tubing was not original to Architect, but installed during a service visit that occurred between June 16, 2017 and July 20, 2017. Peristaltic Head tubing is used on Architect clinical chemistry instruments as part of waste aspiration during the cuvette cleaning process.

Affected products

A. ARCHITECT C 16000 - PROCESSING MODULE

Lot or serial number
  • s/n: C1600144
  • s/n: C1600819
  • s/n: C1600859
  • s/n: C1600976
Model or catalog number

03L77-01

Companies
Manufacturer

Abbott Laboratories Diagnostics Division

100 ABBOTT PARK ROAD

ABBOTT PARK

60064

Illinois

UNITED STATES



B. ARCHITECT C4000 SYSTEM - PROCESSING MODULE

Lot or serial number
  • s/n: C460943
  • s/n: C461550
Model or catalog number

02P24-01

Companies
Manufacturer

Abbott Laboratories Diagnostics Division

100 ABBOTT PARK ROAD

ABBOTT PARK

60064

Illinois

UNITED STATES