This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

AQUARIUS Workstation Basic Unit (2015-01-21)

Starting date:
January 21, 2015
Posting date:
March 5, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-52449

Recalled Products

AQUARIUS Workstation Basic Unit

Reason

Software anomaly related to the Recist 1.1 target lesion evaluation criteria in the findings workflow module within the aquarius workstation intuition client viewer. In certain instances, this target lesion evaluation output provided by the software may be inaccurate. The total measurement calculation and percentage of change are accurate. The affected versions are 4.4.11.82.6784, 4.4.11.116.7134 and 4.4.11.144.7589.

Affected products

AQUARIUS Workstation Basic Unit

Lot or serial number

AQ-UH4940

AQ-UH7915

AQ-UH8040

AQ-UH8486

Model or catalog number

84-0069

Companies
Manufacturer

TeraRecon, Inc.

4000 East 3rd Avenue
Suite 200

Foster City

94404

California

UNITED STATES