Aquac Uno H Reverse Osmosis System - Main Unit
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Aquac Uno H Reverse Osmosis System - Main Unit |
9CHS2905 |
F00004561 |
Issue
The reason for this recall is to inform the operators and users of one Aquac Uno H Reverse Osmosis System of a field safety corrective action due to the possibility of missing ferrite cores in the internal network cable of the device. To meet the requirements of IEC 60601-1 and IEC 60601-1-2 standards, two ferrite cores must be attached to the network cable and the cable routing of the internal network cable must be positioned correctly.
Recall start date: November 14, 2022
Additional information
Details
Fresenius Medical Care AG & Co. KGAA
Else-Kroner-Strasse 1, Bad Homburg, Germany, 61352
Get notified
Receive emails about new and updated recall and safety alerts.