Apollo - Main Unit (2019-10-30)
- Starting date:
- October 30, 2019
- Posting date:
- November 15, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-71615
Last updated: 2019-11-15
Affected Products
Apollo - Main Unit
Reason
Following information received from field, it was found that the number of tilting cycles for some units used in high-workload environments had to be increased with respect to the initial specifications. After laboratory testing of the current angulation chain, it was found that it is very close to the design limit for metal fatigue and, under severe workload conditions might fail before the expected lifetime of 10 years.
Affected products
Apollo - Main Unit
Lot or serial number
s/n 08070501
Model or catalog number
- 9784120231
Companies
- Manufacturer
-
Villa Sistemi Medicali S.P.A.
via Delle Azalee 3
Buccinasco
20090
ITALY