Anti-Human Globulin Anti-IgG (Rabbit) MTS Anti-IgG Card

Starting date:
January 31, 2017
Posting date:
February 28, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62400

Reason

Ortho Clinical Diagnostics (Ortho) has initiated a recall due to the potential for false positive reactions when using specific lots of anti-human globulin Anti-IgG (rabbit) MTS Anti-IgG cards (MTS Anti-IgG cards).

It has been determined that affected MTS Anti-igg card lots may exhibit a varying low-level activity of anti-complement that may result in a false positive test.

Affected products

Anti-Human Globulin Anti-IgG (Rabbit) MTS Anti-IgG Card

Lot or serial number

  • 101216001-50
  • 101216001-51
  • 101216001-53
  • 101216001-55
  • 101216001-56
  • 101216001-57
  • 101216001-58
  • 101216001-60

Model or catalog number

MTS084024

Companies

Manufacturer

Micro Typing Systems Inc.

1295 29th Ave. S.W.

Pompano Beach

33069

Florida

UNITED STATES