This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Anti-HAV Kit

Starting date:
February 9, 2017
Posting date:
February 28, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-62406

Reason

When Li-Heparin and Na-heparin plasma specimens were compared to serum samples during internal investigations, the specified required recovery could not be achieved. Therefore, the claim in the method sheet regarding acceptable specimens under 'specimen collection and preparation' will be corrected.

Affected products

Anti-HAV Kit

Lot or serial number

All lots

Model or catalog number

04854977190

Companies

Manufacturer

Roche Diagnostics GMBH

Sandhoferstrasse 116

Mannheim

68305

GERMANY