Health product recall

Ameriwater MRO Dialysis RO+ System

Starting date:
July 13, 2012
Posting date:
September 17, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-23887

Recalled Products

A. Ameriwater MRO Dialysis RO+ System

Reason

Ameriwater has identified a potential problem with the Ameriwater Portable RO+ System Model MROS20, MROS21, MRO120 and MRO121 with serial numbers between 31000 and 31079. Ameriwater specified 250 Volt 15 Amp fuses, but the supplier of the motor starter mistakenly installed 32 Volt fuses in the assembly.

Affected products

A. Ameriwater MRO Dialysis RO+ System

Lot or serial number

31000-31079

Model or catalog number

00MRO120, 00MRO121, 00MROS20, 00MROS21, MRO1, MROS

Companies
Manufacturer
Ameriwater