Ameriwater MRO Dialysis RO+ System
- Starting date:
- July 13, 2012
- Posting date:
- September 17, 2012
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-23887
Recalled Products
A. Ameriwater MRO Dialysis RO+ System
Reason
Ameriwater has identified a potential problem with the Ameriwater Portable RO+ System Model MROS20, MROS21, MRO120 and MRO121 with serial numbers between 31000 and 31079. Ameriwater specified 250 Volt 15 Amp fuses, but the supplier of the motor starter mistakenly installed 32 Volt fuses in the assembly.
Affected products
A. Ameriwater MRO Dialysis RO+ System
Lot or serial number
31000-31079
Model or catalog number
00MRO120, 00MRO121, 00MROS20, 00MROS21, MRO1, MROS
Companies
- Manufacturer
- Ameriwater