Health product recall

Alphenix Systems

Last updated

Summary

Product
Alphenix Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Alphenix INFX-8000c-Main Unit

S/N: PFD2492006

INFX-8000C/PF

Alphenix INFX-8000c-Main Unit

S/N: QWB2472001

INFX-8000C/QW

Alphenix INFX-8000c-Main Unit

S/N: QWB2512002

INFX-8000C/QW

Alphenix INFX-8000c-Main Unit

S/N: QWB2567001

INFX-8000C/QW

Alphenix INFX-8000c-Main Unit

S/N: QEA2512001

INFX-8000C/QE

Alphenix INFX-8000c-Main Unit

S/N: QUB2482003

INFX-8000C/QU

Alphenix INFX-8000c-Main Unit

S/N: QUB2492004

INFX-8000C/QU

Alphenix INFX-8000c-Main Unit

S/N: QUB2472002

INFX-8000C/QU

Alphenix INFX-8000c-Main Unit

S/N: QUB2557003

INFX-8000C/QU

Alphenix INFX-8000c-Main Unit

S/N: QUB2512005

INFX-8000C/QU

Alphenix INFX-8000v-Main Unit

S/N: Q1B2472001

INFX-8000V/Q1

Alphenix INFX-8000v-Main Unit

S/N: PUB2227001

INFX-8000V/PU

Alphenix INFX-8000v-Main Unit

S/N: PVD2292001

INFX-8000V/PV

Alphenix INFX-8000v-Main Unit

S/N: P5D2492002

INFX-8000V/P5

Alphenix INFX-8000v-Main Unit

S/N: QDB2547002

INFX-8000V/QD

Issue

An internal error would occur if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. The manufacturer confirmed that this is caused by a software problem and will not occur if F-STORE or image acquisition are performed after terminating CAAS clinical analysis function.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Canon Medical Systems Corporation

1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-81590

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe