Alphenix Systems
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Alphenix INFX-8000c-Main Unit |
S/N: PFD2492006 |
INFX-8000C/PF |
Alphenix INFX-8000c-Main Unit |
S/N: QWB2472001 |
INFX-8000C/QW |
Alphenix INFX-8000c-Main Unit |
S/N: QWB2512002 |
INFX-8000C/QW |
Alphenix INFX-8000c-Main Unit |
S/N: QWB2567001 |
INFX-8000C/QW |
Alphenix INFX-8000c-Main Unit |
S/N: QEA2512001 |
INFX-8000C/QE |
Alphenix INFX-8000c-Main Unit |
S/N: QUB2482003 |
INFX-8000C/QU |
Alphenix INFX-8000c-Main Unit |
S/N: QUB2492004 |
INFX-8000C/QU |
Alphenix INFX-8000c-Main Unit |
S/N: QUB2472002 |
INFX-8000C/QU |
Alphenix INFX-8000c-Main Unit |
S/N: QUB2557003 |
INFX-8000C/QU |
Alphenix INFX-8000c-Main Unit |
S/N: QUB2512005 |
INFX-8000C/QU |
Alphenix INFX-8000v-Main Unit |
S/N: Q1B2472001 |
INFX-8000V/Q1 |
Alphenix INFX-8000v-Main Unit |
S/N: PUB2227001 |
INFX-8000V/PU |
Alphenix INFX-8000v-Main Unit |
S/N: PVD2292001 |
INFX-8000V/PV |
Alphenix INFX-8000v-Main Unit |
S/N: P5D2492002 |
INFX-8000V/P5 |
Alphenix INFX-8000v-Main Unit |
S/N: QDB2547002 |
INFX-8000V/QD |
Issue
An internal error would occur if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. The manufacturer confirmed that this is caused by a software problem and will not occur if F-STORE or image acquisition are performed after terminating CAAS clinical analysis function.
Additional information
Details
Canon Medical Systems Corporation
1385, Shimoishigami, Otawara-Shi, Tochigi-Ken, Japan, 324-8550