Allura Xper Systems
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalogue number |
|---|---|---|
Allura Xper FD20 (Product Of AlluraClarity Family) |
Not applicable. |
722028 |
Allura Xper FD 20 Or Table (Product Of AlluraClarity Family) |
Not applicable. |
722035 |
Allura Xper FD10/10 (Product Of AlluraClarity Family) |
Not applicable. |
722027 |
Allura Xper FD10 (Product Of AlluraClarity Family) |
Not applicable. |
722026 |
Allura Xper FD20/20 (Product Of AlluraClarity Family) |
Not applicable. |
722038 |
Allura Xper FD10 |
Not applicable. |
001443 |
Allura Xper FD20/15 |
Not applicable. |
722058 |
Issue
Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.