Health product recall

Allura Xper Systems

Last updated

Summary

Product
Allura Xper Systems
Issue
Medical devices - Performance
What to do

Contact the manufacturer if you require additional information. 

Audience
Healthcare

Affected products

Affected products

Lot or serial number

Model or catalogue number

Allura Xper FD20 (Product Of AlluraClarity Family)

Not applicable.

722028

Allura Xper FD 20 Or Table (Product Of AlluraClarity Family)

Not applicable.

722035

Allura Xper FD10/10 (Product Of AlluraClarity Family)

Not applicable.

722027

Allura Xper FD10 (Product Of AlluraClarity Family)

Not applicable.

722026

Allura Xper FD20/20 (Product Of AlluraClarity Family)

Not applicable.

722038

Allura Xper FD10

Not applicable.

001443

Allura Xper FD20/15

Not applicable.

722058

Issue

Philips has identified that the BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the Allura System will be halted and the Allura System will not start. No user messages are shown before the battery is low on power or depleted.

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Radiology
Companies

Philips Medical Systems Nederland B.V.

Veenpluis 6, Best, Netherlands, 5684 PC

Published by
Health Canada
Audience
Healthcare
Recall class
Type II
Recall date
Identification number
RA-78125

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe