Health product recall

ALLURA XPER FD10/10 - Main Unit, FD20/10 System-Main Unit, FD20/20, FD10, FD20 System - Main Unit, FD20 Or Table (2017-12-28)

Starting date:
December 28, 2017
Posting date:
January 29, 2018
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-65802

Affected products

  1. ALLURA XPER FD10/10 - MAIN UNIT
  2. ALLURA XPER FD20/10 SYSTEM-MAIN UNIT
  3. ALLURA XPER FD20/20
  4. ALLURA XPER FD10
  5. ALLURA XPER FD20 SYSTEM - MAIN UNIT
  6. ALLURA XPER FD20 OR TABLE

Reason

Philips has become aware that extra monitors, so called 7th or 8th monitor, mounted on the Monitor Ceiling Suspension (MCS), might disengage from the MCS and fall down. The two small set screws securing the two gliding rods into an aluminum block of the support arm can come loose because of vibrations and forces exerted on them.

If the set screws have come loose gravity can keep the rods in position. However, if a force is exerted on the monitor pushing it forward, the monitor can topple forward. Once the monitor has toppled over it is only attached with its wires. If the monitor breaks loose of its wires it will drop on the floor.

Affected products

A. ALLURA XPER FD10/10 - MAIN UNIT

Lot or serial number

Not applicable

Model or catalog number

ALLURA XPER FD10/10

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS



B. ALLURA XPER FD20/10 SYSTEM-MAIN UNIT

Lot or serial number

Not applicable

Model or catalog number

ALLURA XPER FD20/10

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS



C. ALLURA XPER FD20/20

Lot or serial number

Not applicable

Model or catalog number

ALLURA XPER FD20/20

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS



D. ALLURA XPER FD10

Lot or serial number

Not applicable

Model or catalog number

ALLURA XPER FD10

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS



E. ALLURA XPER FD20 SYSTEM - MAIN UNIT

Lot or serial number

Not applicable

Model or catalog number

ALLURA XPER FD20

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS



F. ALLURA XPER FD20 OR TABLE

Lot or serial number

Not applicable

Model or catalog number

000454

Companies
Manufacturer

Philips Medical Systems Nederland B.V.

Veenpluis 4-6

Best

5684 PC

NETHERLANDS