ALLURA XPER FD10 and FD20/10 (PRODUCT OF ALLURACLARITY FAMILY)
Summary
Contact the manufacturer if you require additional information
Affected products
Affected Products |
Lot or serial number |
Model or catalog number |
---|---|---|
ALLURA XPER FD10 |
Not applicable |
001443 |
ALLURA XPER FD20/10 (PRODUCT OF ALLURACLARITY FAMILY) |
Not applicable |
722029 |
Issue
A problem has been identified in the Philips Allura Xper and Azurion systems that could pose a risk for patients. Philips has discovered that the wireless foot switch can suddenly stop responding due to a firmware bug when a number of ambient conditions coexist, such as EMC disturbance and the presence of other wireless devices in the room. The wireless footswitch is an option to Philips Allura Xper and Azurion systems to release live Fluoroscopy and exposure. Recovery of this error can be done by means of a cold restart of the system [power down, power up], which restores the functionality of the base station.
Recall start date: 2021-10-25
Additional information
Details
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V
VEENPLUIS 6, BEST
Netherlands, 5684 PC
Get notified
Receive notifications for new and updated recalls and alerts by category.