Allset+T Gold HLA ABC Low Resolution (2014-03-11)

Starting date:
March 11, 2014
Posting date:
April 4, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38683

Recalled products

  1. Allset+T Gold HLA ABC Low Resolution

Reason

When using 54340D Allset+ Gold HLA ABC Low Resolution kit lot number 038 1249868, Primer mix PM085C has been shown to produce a weak false positive reaction of correct size (420 base pair PCR Amplicon) in the presence of samples containing C*04 homozygote allele(s). Primer mix PM085C can be found in lane 78. As a result of the false positive reaction, samples containing two C*04 alleles could produce a mistype or no-type result depending on the reaction pattern corresponding to the full ABC HLA typing.

Affected products

A. Allset+T Gold HLA ABC Low Resolution

Lot or serial number
  • 038 1249868
Model or catalog number
  • 54340D