Health product recall

ALLSET GOLD SSP (2015-04-15)

Starting date:
April 15, 2015
Posting date:
May 8, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53287

Recalled Products

A. ALLSET GOLD DPB1 HIGH RES SSP

B. DPB1 SSP UNITRAY W/TAQ 12 TEST, DPB1 SSP UNITRAY 12 TESTS

Reason

One complaint regarding this issue has been received. One customer complaint (PR184220) was recorded. A customer reported that the reactivity for lanes 13, 27, 30, 39 and 40 was negative for the DPB1*99:01 allele when the labeling of the DPB1 SSP Unitray 12 tests kit (451606D) indicates the reactivity should be positive. The labeling discrepancy may lead to potentially mistyping a DPB1*99:01 allele (population frequency < 0.01%) as a DPB1*39:01 allele. After investigation, it was determined that the DPB1 SSP Unitray with Taq 12 tests kit (451616D) and the Allset Gold DPB1 High RES SSP kit (54070D) are also affected by the labeling issue since they contain the same primer present in the DPB1 SSP Unitray 12 tests kit (451606d). Life technologies confirmed the issue on 11 March 2015.

Affected products

A. ALLSET GOLD DPB1 HIGH RES SSP

Lot or serial number

009 1377591

009 1450550

009 1544263

009 1608835

009 1618548

Model or catalog number

54070D

Companies
Manufacturer

Life Technologies Corporation

9099 North Deerbrook Trail

Brown Deer

53223

Wisconsin

UNITED STATES



B. DPB1 SSP UNITRAY W/TAQ 12 TEST, DPB1 SSP UNITRAY 12 TESTS

Lot or serial number

More than 10 numbers, contact manufacturer.

Model or catalog number

451606D

451616D

Companies
Manufacturer

Life Technologies Corporation

9099 North Deerbrook Trail

Brown Deer

53223

Wisconsin

UNITED STATES