Health product recall

Allset Gold DPB1 High Resolution SSP Kit, DPB1 SSP UniTray 12 Tests and DPB1 SSP UniTray With Taq 12 Tests (2014-08-14)

Starting date:
August 14, 2014
Posting date:
September 16, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-41365

Recalled products

  1. Allset Gold DPB1 High Resolution SSP Kit
  2. DPB1 SSP UniTray 12 Tests and DPB1 SSP UniTray With Taq 12 Tests

Reason

On June 26, 2014, one customer complaint was received for this issue (internal ref. # PR162312): a customer reported that a sample had discrepant results when using SSP UniTray versus two other typing methods. SSP UniTray typed the sample as DPB1*04:01, DPB1*81:01. The correct assignment should be DPB1*04:01, DPB1*33:01. This is considered a mistype result.

Affected products

A. Allset Gold DPB1 High Resolution SSP Kit

Lot or serial number
  • 008 1204422
  • 008 1256851
  • 009 1377591
  • 009 1450550
  • 009 1544263
Model or catalog number
  • 54070D
Companies
Manufacturer

Life Technologies Corporation (Invitrogen)

9099 North Deerbrook Trail

Brown Deer

53223

Wisconsin

UNITED STATES



B. DPB1 SSP UniTray 12 Tests and DPB1 SSP UniTray With Taq 12 Tests

Lot or serial number
  • 008 1154360
  • 008 1190373
  • 008 1220578
  • 008 1289713
  • 009 1415057
  • 009 1464121
  • 009 1568516
Model or catalog number
  • 451606D
  • 451616D
Companies
Manufacturer

Life Technologies Corporation (Invitrogen)

9099 North Deerbrook Trail

Brown Deer

53223

Wisconsin

UNITED STATES