Alissa Interpret
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
Alissa Interpret |
v5.3.7 |
K5849AA |
Issue
1. The ACMG (American College of Medical Genetics and Genomics) classification scheme implemented in Alissa Interpret does not result in variant of uncertain significance (VOUS) classification when benign and pathogenic variant are contradictory.
2. The ACMG classification scheme does not result in the correct classification when the criteria strength is updated.
Recall Start Date: December 21, 2023
Additional information
Details
Agilent Technologies Belgium
Technologielaan 3, Leuven, Belgium,
Get notified