Alinity i Processing Module (2019-04-09)
- Starting date:
- April 9, 2019
- Posting date:
- April 18, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-69652
Last updated: 2019-04-18
Affected Products
Alinity i Processing Module
Reason
There is a potential for loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. A loose cable connection has the potential to heat up and cause damage to the wire and connector. If this occurs, the operator may detect a burning smell. The cable connection to the reagent cooler is located within the processing module and is not intended to be accessed by the operator. Due to the very low flammability rating of the materials used, any heat damage would be localized to the connectors. The cable connections on the reagent cooler have been redesigned to include more robust connectors.
Affected products
Alinity i Processing Module
Lot or serial number
- AI01053
- AI01072
- AI01112
- AI01133
- AI01242
Model or catalog number
03R65-01
Companies
- Manufacturer
-
Abbott GmbH & Co. KG
Max-Planck-Ring 2
Wiesbaden
65205
GERMANY