ALINITY I PROCESSING MODULE (2017-10-13)
- Starting date:
- October 13, 2017
- Posting date:
- October 27, 2017
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- Healthcare Professionals, General Public, Hospitals
- Identification number:
- RA-64958
Affected products
ALINITY I PROCESSING MODULE
Reason
Abbott has identified the potential for a loose cable connection on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. The loose cable connection has the potential to heat up over long periods of time and cause damage to the wire and connector. This would be detected by the operator in the form of a burning smell. The cable connection to the reagent cooler is located within the processing module and is not accessed by the operator. Due to the very low flammability rating of the materials used, the heat damage is localized to the connector.
Affected products
ALINITY I PROCESSING MODULE
Lot or serial number
AI01072
Model or catalog number
03R65-01
Companies
- Manufacturer
-
Abbott GmbH and Co. K.G.
Max Planck Ring 2
Wiesbaden
65205
GERMANY