Health product recall

ALINITY I PROCESSING MODULE (2017-10-13)

Starting date:
October 13, 2017
Posting date:
October 27, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-64958

Affected products

ALINITY I PROCESSING MODULE

Reason

Abbott has identified the potential for a loose cable connection on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. The loose cable connection has the potential to heat up over long periods of time and cause damage to the wire and connector. This would be detected by the operator in the form of a burning smell. The cable connection to the reagent cooler is located within the processing module and is not accessed by the operator. Due to the very low flammability rating of the materials used, the heat damage is localized to the connector.