Alinity ci-series System Control Module (2019-06-13)
- Starting date:
- June 13, 2019
- Posting date:
- June 28, 2019
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type II
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-70321
Last updated: 2019-06-28
Affected Products
Alinity ci-series System Control Module
Reason
Abbott has identified an issue with all on-market versions of Alinity ci-series Software where re-use of reaction vessels (RVs) may occur after a system Stop. This issue only occurs if the system is transitioned from Processing to Stopped to Idle and has the potential to cause incorrect results.
Affected products
Alinity ci-series System Control Module
Lot or serial number
All lots.
Model or catalog number
03R70-01
Companies
- Manufacturer
-
Abbott GMBH & Co. K.G.
MAX PLANCK RING 2
WIESBADEN
65205
GERMANY