This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ALB2 (Albumin Gen. 2) for Cobas C701 Analyzer

Starting date:
October 3, 2012
Posting date:
October 29, 2012
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public (GP), Healthcare Professionals, Hospitals
Identification number:
RA-15163

Recalled Products:

ALB2 (Albumin Gen. 2) for Cobas C701 Analyzer

Reason

Is related to a possible issue where patient and quality control albumin results may become erratic over time with a continuous throughput of more than 300 tests in a shift with deviations of up to 10%.

Affected products

ALB2 (Albumin Gen. 2) for Cobas C701 Analyzer

Lot or serial number

All lots

Model or catalog number

05166861190

Companies
Manufacturer

Troche Diagnostics Ltd.