AISYS CS2 ANESTHESIA MACHINE MAIN UNIT (2016-02-18)

Starting date:
February 18, 2016
Posting date:
March 4, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
Healthcare Professionals, General Public, Hospitals
Identification number:
RA-57396

Affected Products

A. AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Reason

GE Healthcare has recently become aware of a potential safety issue involving a momentary, self-correcting anesthetic agent bolus when using 21% oxygen on all Aisys CS2 and upgraded Aisys anesthesia devices. Delivery of this momentary bolus of anesthetic agent is potentially hazardous because it could lead to hypotension in certain vulnerable pediatric patients when 21% oxygen (air only) is used. There have been no injuries reported as a result of this issue.

Affected products

A. AISYS CS2 ANESTHESIA MACHINE MAIN UNIT

Lot or serial number

All Lots

Model or catalog number

1011-9050-000

Companies
Manufacturer
Datex-Ohmeda
3030 Ohmeda Drive, P.O. Box 7550,
Madison
53707-7550
Wisconsin
UNITED STATES