ADVIA® and Atellica® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay
Brand(s)
Summary
Contact the manufacturer if you require additional information.
Affected products
Affected products |
Lot or serial number |
Model or catalog number |
|---|---|---|
ADVIA® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay |
140 |
11319151 |
Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) |
130414 |
Not applicable. |
Issue
Siemens Healthineers has confirmed the potential for biased quality control (qc) and patient results when using Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) lot 130414 on the Stellica® CH and Stellica® CI analyzers. Siemens' internal investigation confirmed when using lot 130414, QC may recover outside of the allowable control limits for urine chemistry and spinal fluid control levels.
Recall start date: August 1, 2024
Additional information
Details
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave, Tarrytown, New York, United States, 10591