Health product recall

ADVIA® and Atellica® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay

Last updated

Summary

Product
ADVIA® and Atellica® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay
Issue
Medical devices - Device compatibility
What to do

Contact the manufacturer if you require additional information. 

Affected products

Affected products Lot or serial number Model or catalog number
ADVIA® Chemistry Urinary/Cerebrospinal Fluid Protein (UCFP) Assay 140 11319151 11097543
Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) 130414 Not applicable.

Issue

Siemens Healthineers has confirmed the potential for biased quality control (qc) and patient results when using  Atellica® CH Urinary/Cerebrospinal Fluid Protein (UCFP) lot 130414 on the Stellica® CH and Stellica® CI analyzers. Siemens' internal investigation confirmed when using lot 130414, QC may recover outside of the allowable control limits for urine chemistry and spinal fluid control levels.

Recall start date: August 1, 2024

Additional information

Details
Original published date:
Alert / recall type
Health product recall
Category
Health products - Medical devices - Anaesthesiology
Health products - Medical devices - Chemistry
Companies

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave, Tarrytown, New York, United States, 10591

Published by
Health Canada
Audience
General public
Healthcare
Industry
Recall class
Type II
Identification number
RA-75952

Get notified

Receive emails about new and updated recall and safety alerts.

Subscribe