Health product recall

ADVIA CHEMISTRY SYSTEMS REAGENTS (2015-09-24)

Starting date:
September 24, 2015
Posting date:
October 14, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type II
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-55360

Products Affected

  1. ADVIA 1200 CHEMISTRY SYSTEM-HEMOGLOBIN A1C_3 REAGENTS
  2. ADVIA CHEMISTRY SYSTEMS-HEMOGLOBIN A1C_3 REAGENTS

Reason

Siemens Healthcare Diagnostics has confirmed the Hemoglobin A1c_3 and A1c_3M reagent affected kit lots used on the ADVIA 1200, 1650, 1800, 2400, and XPT Chemistry Systems may demonstrate an increased occurrence of high %HbA1c (glycosylated hemoglobin) bias. Siemens internal investigation demonstrates reagent lots 230 and 231 may exhibit a positive bias averaging 0.6% HbA1c units, ranging from -0.1% to 1.1% HbA1c units. The bias was observed when comparing %HbA1c means to National Glycohemoglobin Standardization Program (NGSP) pooled patient target-value assigned samples ranging from approximately 5.5% to 8.0% HbA1c. The maximum bias was observed at higher %HbA1c concentrations. Quality control samples may exhibit a similar bias.

Affected products

A. ADVIA 1200 CHEMISTRY SYSTEM-HEMOGLOBIN A1C_3 REAGENTS

Lot or serial number

LOT 230

Model or catalog number

10485591

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES



B. ADVIA CHEMISTRY SYSTEMS-HEMOGLOBIN A1C_3 REAGENTS

Lot or serial number

LOT 230

Model or catalog number

10485591

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New York

UNITED STATES