ADVIA CHEMISTRY SYSTEMS (2016-10-07)
- Starting date:
- October 7, 2016
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-60718
Affected Products
A, ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS
B. ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS
Reason
Siemens has confirmed that the DBIL_2 assigned calibrator values (ACV) currently assigned to certain lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for direct bilirubin measurement. Due to the proportional bias, the differences become greater as the concentration of DBIL_2 increases. Quality control (QC) material is also affected by this issue. Depending on the quality control concentrations and ranges used by the laboratory, quality control may not detect this issue.
Affected products
A, ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS
Lot or serial number
680726
680726A
680726B
680726C
680726D
Model or catalog number
10312279
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New Jersey
UNITED STATES
B. ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS
Lot or serial number
680726
680726A
680726B
680726C
680726D
Model or catalog number
10312279
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
10591
New Jersey
UNITED STATES