Health product recall

ADVIA CHEMISTRY SYSTEMS (2016-10-07)

Starting date:
October 7, 2016
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-60718

Affected Products

A, ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS

B. ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS

Reason

Siemens has confirmed that the DBIL_2 assigned calibrator values (ACV) currently assigned to certain lots may cause, on average, a positive proportional bias of approximately 41% on patient samples compared to the reference method for direct bilirubin measurement. Due to the proportional bias, the differences become greater as the concentration of DBIL_2 increases. Quality control (QC) material is also affected by this issue. Depending on the quality control concentrations and ranges used by the laboratory, quality control may not detect this issue.

Affected products

A, ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS

Lot or serial number

680726

680726A

680726B

680726C

680726D

Model or catalog number

10312279

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New Jersey

UNITED STATES



B. ADVIA CHEMISTRY SYSTEMS - CHEMISTRY CALIBRATORS

Lot or serial number

680726

680726A

680726B

680726C

680726D

Model or catalog number

10312279

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

10591

New Jersey

UNITED STATES