This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

Advia Chemistry Systems (2014-01-23)

Starting date:
January 23, 2014
Posting date:
March 19, 2014
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-38305

Recalled Products

Advia Chemistry Systems 

Reason

Siemens Healthcare Diagnostics is conducting a recall for Advia 1800 and Advia 2400 ISE firmware v2.23. Advia Chemistry Chloride Electrodes may fail prematurely during calibration due to a chloride bias parameter error. The operator will receive an ng flag which requires immediate action to be taken on the calibration samples.

Affected products

A. Advia Chemistry System - Advia 1800 Analyzer Class 2

Lot or serial number

All lots

Model or catalog number

01412173

073-A021-01

10315183

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES



B. Advia Chemistry System - Advia 2400 Analyzer Class 2

Lot or serial number

All lots

Model or catalog number

01714935

073-A011-02

10315769

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave.

Tarrytown

10591

New York

UNITED STATES