This page has been archived on the Web
Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.
Advia Chemistry System Products
- Starting date:
- January 16, 2013
- Posting date:
- February 4, 2013
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-20247
Recalled Products
- Advia Chemistry System – Alkaline Phosphatase (ALPAMP) Assay
- Advia 1200 Chemistry System – Alkaline Phosphatase (ALPAMP) Assay
Reason
There is an increased incidence of absorbance flags (u, u, or ////) on some ALPAMP and ALPDEA reagent wedges, which in some cases can impede reporting of results. No QC shift or discrepant patient results are observed.
Affected products
A. Advia Chemistry System – Alkaline Phosphatase (ALPAMP) Assay
Lot or serial number
- 255406
- 260121
Model or catalog number
03035814
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
B. Advia 1200 Chemistry System – Alkaline Phosphatase (ALPAMP) Assay
Lot or serial number
- 255406
- 260121
Model or catalog number
03035814
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.