ADVIA Chemistry System - High Sensitivity C-Reactive Protein(HSCRP) Assay (2019-11-21)

Starting date:
November 21, 2019
Posting date:
January 6, 2020
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-71989

Last updated: 2020-01-07

Affected Products

ADVIA Chemistry System - High Sensitivity C-Reactive Protein(HSCRP) Assay

Reason

Recall due to current assigned values for ADVIA Chemistry HSCRP Calibrator lot 453610 demonstrate a positive bias of approximately 11% for HSCRP patient samples and quality control material when compared to the European reference material ERM da474/IFCC for CRP measurement.

Affected products

ADVIA Chemistry System - High Sensitivity C-Reactive Protein(HSCRP) Assay

Lot or serial number

453610

Model or catalog number