This page has been archived on the Web

Information identified as archived is provided for reference, research or recordkeeping purposes. It is not subject to the Government of Canada Web Standards and has not been altered or updated since it was archived. Please contact us to request a format other than those available.

Health product recall

ADVIA CHEMISTRY SYSTEM (2015-04-13)

Starting date:
April 13, 2015
Posting date:
May 6, 2015
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-53245

Recalled Products

A. ADVIA CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY

B. ADVIA 1200 CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY

Reason

Amylase reagent kit for use on the ADVIA Chemistry Systems demonstrates an increased incidence of absorbance flags (U or U accompanied by U). Absorbance flags (U) indicate an abnormally high reagent blank absorbance exceeding the reagent blank limit. In some cases with these reagent lots, sample results may also have a sample absorbance (U) flag indicating a high absorbance for the sample that occurs in conjunction with the (U) flag. Amylase results can be generated with these flags.

Affected products

A. ADVIA CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY

Lot or serial number

304821

Model or catalog number

07498401

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

New York

UNITED STATES



B. ADVIA 1200 CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY

Lot or serial number

304816

Model or catalog number

3031177

Companies
Manufacturer

Siemens Healthcare Diagnostics Inc.

511 Benedict Ave

Tarrytown

New York

UNITED STATES