ADVIA CHEMISTRY SYSTEM (2015-04-13)
- Starting date:
- April 13, 2015
- Posting date:
- May 6, 2015
- Type of communication:
- Medical Device Recall
- Subcategory:
- Medical Device
- Hazard classification:
- Type III
- Source of recall:
- Health Canada
- Issue:
- Medical Devices
- Audience:
- General Public, Healthcare Professionals, Hospitals
- Identification number:
- RA-53245
Recalled Products
A. ADVIA CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY
B. ADVIA 1200 CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY
Reason
Amylase reagent kit for use on the ADVIA Chemistry Systems demonstrates an increased incidence of absorbance flags (U or U accompanied by U). Absorbance flags (U) indicate an abnormally high reagent blank absorbance exceeding the reagent blank limit. In some cases with these reagent lots, sample results may also have a sample absorbance (U) flag indicating a high absorbance for the sample that occurs in conjunction with the (U) flag. Amylase results can be generated with these flags.
Affected products
A. ADVIA CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY
Lot or serial number
304821
Model or catalog number
07498401
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
New York
UNITED STATES
B. ADVIA 1200 CHEMISTRY SYSTEM - AMYLASE (AMYLAS) ASSAY
Lot or serial number
304816
Model or catalog number
3031177
Companies
- Manufacturer
-
Siemens Healthcare Diagnostics Inc.
511 Benedict Ave
Tarrytown
New York
UNITED STATES