Health product recall

Advia Chemistry Enzyme 2 Calibrator

Starting date:
July 18, 2017
Posting date:
August 15, 2017
Type of communication:
Medical Device Recall
Subcategory:
Medical Device
Hazard classification:
Type III
Source of recall:
Health Canada
Issue:
Medical Devices
Audience:
General Public, Healthcare Professionals, Hospitals
Identification number:
RA-64212

Reason

Siemens Healthcare Diagnostics has confirmed a positive bias of up to 41% in quality control and patient results for aspartate aminotransferase (AST) and aspartate aminotransferase concentrated (AST_C) assays following calibration with Advia Chemistry Enzyme 2 Calibrator Lots 6JD079 and 6MD042. Only aspartate aminotransferase (AST) and aspartate aminotransferase concentrated (AST_C) assays are affected. All other assays (ALT, ALT_C, ALTP5P, ALTP_C, ASTP5P and ASTP_C) using the Advia Chemistry Enzyme 2 Calibrator are not impacted.

Affected products

Advia Chemistry Enzyme 2 Calibrator

Lot or serial number

6JD079

6MD042

Model or catalog number

10916058